SOFTCONTROL V-UE BLEND
For certain processes, a low water hardness is required. To achieve this, the soft water is blended with the raw water. The continuous monitoring of the water hardness behind a blending system with Softcontrol-V / UE is thus of great importance.
In contrast to conventional water monitoring devices Softcontrol-V/UE operates with a selective sensor for calcium and magnesium ions on potentiometric method basis.
The monitored upper threshold value can be parameterised at the device. With exceeding of this threshold value, alarm will be triggered by a contact. Also an alarm delay is parameterisable individually. Measurement values and status information can be stored on an integrated USB flash drive.
-
Softcontrol-V/UE blend can be installed in every blending system for water.
-
No reagents or indicators necessary, making the device environmental friendly and cost-effective.
-
Except for the occasional sensor exchange Softcontrol V-UE Blend is almost maintenance free.
SOFTCONTROL V-UE BLEND TECHNICAL DATA | |
Wall Mounted | 400 x 250 x 160 mm (W x H x D) |
Weight | Approx 8 kg |
Surrounding Temperature | 5 °C … 50 °C |
Relative Humidity | 20 % … 80 % |
Power Supply | 100 … 240 Volt / 50 … 60 Hz |
Power Consumption | 15 V DC, about 20 Watt |
Raw and Blended Water Connection | Branch pipe with connection for PA-tube o.d. 4 mm, i.d. 2 mm, up streamed dirt filter ≤ 0,1 mm recommended |
Raw and Blended Water Pressure | Minimum 1 bar up to maximum 10 bars |
Drain | Minimum ø 15 mm, pressure free |
Quality of Raw and Soft Water | Natural water, free of grease, oil and brine germ content < 5000 CFU/ml |
Hardness Range Raw Water | 0.72 mmol/l ... 32.4 mmol/l |
Possibility of Calibration Input Actual Value of Blended Water Hardness | 0.2 mmol/l ... 10,8 mmol/l |
Setting of Threshold Value | 0.2 mmol/l ... 10,8 mmol/l |
Sensor Control | Automatic |
External Signaling | Potential-free contacts (sensor alarm, threshold value alarm) 4 … 20 mA current loop, paramterisable |
Operation Time of Sensor | Approx. 6 to 9 months |